Varios 970

QNo / Poor vibration - TS26403

A
The Front Panel does not light, even if the Power Switch is ON.
Cause1
The Power Cord or the Jack is disconnected.
Solution1
Correctly insert the Power Cord or the Jack.
Cause2
The Fuse is burned out.
Solution2
Contact dealer.*
The Tip does not generate vibration, in spite of depressing the Foot Control.
Cause1
The Tip is not tightened firmly.
Solution1
Tighten the Tip until the Tip Wrench clicks.
Cause2
Worn Tip.
Solution2
Replace the Tip.
Cause3
Power has not been correctly adjusted for the Tip.
Solution3
Adjust the power on the Power Guide or Tip case label. Do not exceed.
Cause4
The Foot Control is disconnected.
Solution4
Connect the Foot Control correctly.
Cause5
Failure of vibrator in the handpiece.
Solution5
Contact dealer.*
Cause6
Failure of internal components of the Foot Control.
Solution6
Contact dealer.*
* Repairs cannot be made by the customer.

QThe Tip is bent or broken. - TS26405

A
Cause
Power has not been properly adjusted for the Tip.
Solution
Adjust the power level the Power Guide or Tip case label. Do not exceed.

QThe Tip is flying away. - TS26407

A
Cause
The Tip is not tightened firmly.
Solution
Tighten the Tip until the Tip Wrench clicks.

QNoise from the handpiece. - TS26409

A
Cause1
Power has not been properly adjusted for the Tip.
Solution1
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
Cause2
The Tip is not tightened firmly.
Solution2
Tighten the Tip until the Tip Wrench clicks.
Cause3
Failure of vibration in the handpiece or the Control Unit.
Solution3
Contact dealer.*
* Repairs cannot be made by the customer.

QThe handpiece is overheating. - TS26411

A
Cause1
Power has not been properly adjusted for the Tip.
Solution1
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
Cause2
The Tip is not tightened firmly.
Solution2
Tighten the Tip until the Tip Wrench clicks.
Cause3
Failure of vibration in the handpiece or the Control Unit.
Solution3
Contact dealer.*
* Repairs cannot be made by the customer.

QNo Irrigation supply and/or unstable Irrigation supply (Use of Bottle) - TS26413

A
The Irrigation Pump is running.
Cause
The tube twisted.
Solution
Straighten the twisted Irrigation Tube.
The Irrigation Pump is stopping.
Cause
Time to replace Irrigation Pump. (Approx. 500hours after used.)
Solution
Replace with new Irrigation Pump (Refer to 12. (4) Changing the Irrigation Pump).

QNo / Poor water. (Use of Tap Water) - TS26415

A
The water does not reach to the Control Unit.
Solution
Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm2)
Check to see if water reaches the Control Unit.
Cause1
The Water Adjustment Knob is closed.
Solution1
Turn the Water Adjustment Knob and adjust to the appropriate volume.
Cause2
Disconnected Irrigation supply at low volume range. (less than 10ml/min.)
Solution2
No problem. Turn the Water Adjustment Knob and increase the Irrigation volume.
Cause3
The Water Filter is clogged.
Solution3
Replace with new Water Filter (Refer to 12. (6) Changing Water Filter (Option) ).

QWater leakage. - TS26417

A
Water is leaking from the joint between the Irrigation Tube and the Irrigation Connector.
Cause
The Irrigation Tube is not connected correctly.
Solution
Firmly insert the Irrigation Tube into the Irrigation Connector inmost.
Water is leaking from the joint between the handpiece and the cord.
Cause
O-Ring at the handpiece cord is worn or damaged.
Solution
Replace with new O-Ring (Refer to 12 (3) Changing O-Ring •Handpiece Cord).
Water is leaking from the Control Unit.
Cause
The water circuitry in the Control Unit is damaged.
Solution
Contact dealer.*
* Repairs cannot be made by the customer.

QHandpiece LED does not illuminate. (Varios 970 LUX) - TS26419

A
Tip oscillates, but Handpiece LED turns on and off.
Cause
The handpiece is not connected into the Handpiece Cord correctly.
Solution
Firmly insert the handpiece into the Handpiece Cord inmost.
Tip oscillates, but Handpiece LED does not turn on.
Cause
Disconnection in the Handpiece Cord, or failure in the Control Unit.
Solution
Contact dealer.*
* Repairs cannot be made by the customer.

QStart Beeping - TS26421

A
Beeping while power on.
Cause
Depress Foot Control.
Solution
Release the Foot Control.
Beeping while stopping vibration of Tips.
Cause
Abnormal heating of the Control Unit.
Solution
Stop the operation and leave until Control Unit becomes cool.

Varios 370

QNo / Poor vibration - TS26461

A
With power on and no Power Indicator
Cause
The AC Adaptor or the DC Plug is disconnected
Solution
Correctly insert the AC Adaptor or the DC Plug
The Tip does not generate vibration, in spite of depressing the Foot Control
Cause1
The Tip is not tightened firmly
Solution1
Tighten the Tip until the Tip Wrench clicks
Cause2
Worn Tip
Solution2
Replace the Tip
Cause3
Power has not been correctly adjusted for the Tip
Solution3
According to Power Guide or Tip case label. Do not exceed recommendations.
Cause4
The Foot Control is disconnected
Solution4
Connect the Foot Control correctly
Cause5
Failure of vibrator in the handpiece
Solution5
Contact dealer*
Cause6
Failure of internal components of the Foot Control
Solution6
Contact dealer*
* Repairs cannot be made by the customer.

QThe Tip is bent or broken - TS26463

A
Cause
Power has not been properly adjusted for the Tip
Solution
According to Power Guide or Tip case label. Do not exceed recommendations.

QThe Tip is flying away - TS26465

A
Cause
The Tip is not tightened firmly
Solution
Tighten the Tip until the Tip Wrench clicks

QNoise from the handpiece - TS26467

A
Cause1
Power has not been properly adjusted for the Tip
Solution1
According to Power Guide or Tip case label. Do not exceed recommendations.
Cause2
The Tip is not tightened firmly
Solution2
Tighten the Tip until the Tip Wrench clicks
Cause3
Failure of vibration in the handpiece or the Control Unit
Solution3
Contact dealer*
* Repairs cannot be made by the customer.

QThe handpiece is overheating - TS26472

A
Cause1
Power has not been properly adjusted for the Tip
Solution1
According to Power Guide or Tip case label. Do not exceed recommendations.
Cause2
The Tip is not tightened firmly
Solution2
Tighten the Tip until the Tip Wrench clicks
Cause3
Failure of vibration in the handpiece or the Control Unit
Solution3
Contact dealer*
* Repairs cannot be made by the customer.

QNo / Poor water - TS26474

A
The water does not reach to the Control Unit
Solution
Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm2)
Check to see if water reaches the Control Unit
Cause1
The Water Adjustment Knob is closed.
Solution1
Turn the Water Volume Knob and adjust to the appropriate volume
Cause2
Disconnected water supply at low volume range. (less than 10ml/min.)
Solution2
Adjust the Water Volume Knob and increase the Irrigation volume
Cause3
The Water Filter is clogged
Solution3
Replace with new Water Filter (Refer to 11-3 Changing Water Filter (Option) )

QWater leakage - TS26476

A
Water is leaking from the joint between the handpiece and the cord
Cause
O-Ring at the handpiece cord is worn or damaged
Solution
Replace with new O-Ring (Refer to 11-2 Changing O-Ring Handpiece Cord)

QAttachment of the Control Unit Holder is loose - TS26478

A
Cause
Holder worn
Solution
Contact dealer*
* Repairs cannot be made by the customer.

QHandpiece LED does not illuminate. (Varios2 Lux) - TS26480

A
Tip oscillates, but Handpiece LED turns on and off
Cause
The handpiece is not connected into the Handpiece Cord correctly
Solution
Firmly insert the handpiece into the Handpiece Cord completely

QLoss of the power output without operation - TS26482

A
Power output is set 8 at G
Cause
Protection Circuit is activated
Solution
Power output will weaken automatically while continuous operation is over 10min at the setting of Maximum power at G mode. Releasing the foot from the Foot Control. Decrease the Power less than 5, once then increase the power again. (Refer to 7-5 Protection Circuit)